|
Chapter 6 - Temperature, Salinity, and Density
6.9 Measurement of Temperature and Salinity with Depth
Temperature, salinity, and pressure are measured as a function of depth
using various instruments or techniques, and density is calculated from
the measurements.
Bathythermograph (BT)
A mechanical device that measured temperature vs depth on a smoked glass
slide. The device was widely used to map the thermal structure of the
upper ocean, including the depth of the mixed layer before being replaced
by the expendable bathythermograph in the 1970s.
Expendable Bathythermograph (XBT)
An electronic device that measures temperature vs depth using a thermistor
on a free-falling streamlined weight. The thermistor is connected to an
ohm-meter on the ship by a thin copper wire that is spooled out from the
sinking weight and from the moving ship. The XBT is now the most widely
used instrument for measuring the thermal structure of the upper ocean.
Approximately 65,000 are used each year.
The streamlined weight falls through the water at a constant velocity.
So depth can be calculated from fall time with an accuracy of ±2%.
Temperature accuracy is ±0.1°C. And, vertical resolution is typically
65 cm. Probes reach to depths of 200 m to 1830 m depending on model.
Nansen Bottles
(Figure 6.16) Deployed from ships stopped at hydrographic stations. Hydrographic
stations are places where oceanographers measure water properties from
the surface to some depth, or to the bottom, using instruments lowered
from a ship. Usually 20 bottles were attached at intervals of a few tens
to hundreds of meters to a wire lowered over the side of the ship. The
distribution with depth was selected so that most bottles are in the upper
layers of the water column where the rate of change of temperature in
the vertical is greatest. A protected reversing thermometer for measuring
temperature was attached to each bottle along with an unprotected reversing
thermometer for measuring depth. The bottle contains a tube with valves
on each end to collect sea water at depth. Salinity was determined by
laboratory analysis of water sample collected at depth.
After bottles had been attached to the wire and all had been lowered
to their selected depths, a lead weight was dropped down the wire. The
weight tripped a mechanism on each bottle, and the bottle flipped over,
reversing the thermometers, shutting the valves and trapping water in
the tube, and releasing another weight. When all bottles had been tripped,
the string of bottles was recovered. The deployment and retrieval typically
took several hours.
|
|
Figure 6.16 Left: A
CTD ready to be lowered
over the side of a ship. From Davis (1987). Right: Nansen
water bottles before (I), during (II), and after (III) reversing. Both instruments
are shown at close to the same scale. From Defant (1961: 33).
|
CTD
Mechanical instruments on Nansen bottles were replaced beginning in the
1960s by an electronic instrument, called a CTD, that measured conductivity,
temperature, and depth (Figure 6.16). The measurements are recorded in
digital form either within the instrument as it is lowered from a ship
or on the ship. Temperature is usually measured by a thermistor; conductivity
is measured by induction; pressure is measured by a quartz crystal. Modern
instruments have accuracy summarized in Table 6.2.
Table 6.2 Summary of Measurement Accuracy
Variable |
Range |
Best Accuracy |
Temperature |
42°C |
±0.001°C
|
Salinity |
1 |
±0.02 by titration
|
±0.005 by conductivity
|
Pressure |
10,000 dbar |
± 0.65 dbar
|
Density |
2 kg/m3 |
± 0.005 kg/m3
|
Equation of State |
|
± 0.005 kg/m3
|
chapter contents
|